Ayuda
Ir al contenido

Dialnet


Bone Alignment Using the Iterative Closest Point Algorithm

  • Autores: Maarten Beek, Carolyn F. Small, Randy E. Ellis, Richard W. Sellens, David R. Pichora
  • Localización: Journal of applied biomechanics, ISSN-e 1543-2688, ISSN 1065-8483, Vol. 26, Nº. 4, 2010, págs. 526-530
  • Idioma: inglés
  • Texto completo no disponible (Saber más ...)
  • Resumen
    • Computer assisted surgical interventions and research in joint kinematics rely heavily on the accurate registration of three-dimensional bone surface models reconstructed from various imaging technologies. Anomalous results were seen in a kinematic study of carpal bones using a principal axes alignment approach for the registration. The study was repeated using an iterative closest point algorithm, which is more accurate, but also more demanding to apply. The principal axes method showed errors between 0.35 mm and 0.49 mm for the scaphoid, and between 0.40 mm and 1.22 mm for the pisiform. The iterative closest point method produced errors of less than 0.4 mm. These results show that while the principal axes method approached the accuracy of the iterative closest point algorithm in asymmetrical bones, there were more pronounced errors in bones with some symmetry. Principal axes registration for carpal bones should be avoided.


Fundación Dialnet

Dialnet Plus

  • Más información sobre Dialnet Plus

Opciones de compartir

Opciones de entorno