Ayuda
Ir al contenido

Dialnet


Resumen de Development of a Next-Generation Sequencing Method for BRCA Mutation Screening:: A Comparison between a High-Throughput and a Benchtop Platform

Maurice Chan, Shen Mo Ji, Zhen Xuan Yeo, Linda Gan, Eric Yap, Yoon Sim Yap, Raymond Ng, Puay Hoon Tan, Gay Hui Ho, Peter Ang, Ann Siew Gek Lee

  • In a clinical setting, next-generation sequencing (NGS) approaches for the enrichment and resequencing of DNA targets may have limitations in throughput, cost, or accuracy. We evaluated an NGS workflow for targeted DNA sequencing for mutation detection. Targeted sequence data of the BRCA1 and BRCA2 genes, generated using a PCR-based, multiplexed NGS approach using the SOLiD 4 (n = 24) and Ion Torrent PGM (n = 20) next-generation sequencers, were evaluated against sequence data obtained by Sanger sequencing. The overall sensitivity for SOLiD and PGM were 97.8% (95% CI = 94.7 to 100.0) and 98.9% (95% CI = 96.8 to 100.0) respectively. The specificity for the SOLiD platform was high, at 100.0% (95% CI = 99.3 to 100.0). PGM correctly identified all 3 indels, but 68 false-positive indels were also called. Equimolar normalization of amplicons was not necessary for successful NGS. Both platforms are highly amenable to scale-up, potentially reducing the reagent cost for BRCA testing to


Fundación Dialnet

Dialnet Plus

  • Más información sobre Dialnet Plus