Sebastian Quaas, René Loos, Heike Rudolph, Ralph Luthardt
This study aimed to evaluate the correspondence of intraoral digitization (ID) with extraoral digitization (ED) after impression taking. One-stage putty-and wash impressions and ID were carried out in a randomized order for 10 subjects. The impressions were used to make casts, which were then subjected to ED. ID datasets were aligned to create computer-aided design reference models. Deviations between ID and ED were calculated. The mean positive and negative deviations were 37.7 and –48.4 µm, respectively, for one quadrant. The results showed that the ID system is well suited for the acquisition of single-tooth restorations and is of limited suitability for the acquisition of small multiple unit restorations.