Elastically Suspending the Screw Holes of a Locked Osteosynthesis Plate Can Dampen Impact Loads

  • Autores: Felix Capanni, Kirk Hansen, Steven M. Madey, Michael Bottlang, Daniel C. Fitzpatrick
  • Localización: Journal of applied biomechanics, ISSN-e 1543-2688, ISSN 1065-8483, Vol. 29, Nº. 3, 2013, págs. 164-169
  • Idioma: inglés
  • Texto completo no disponible (Saber más ...)
  • Resumen
    • Impact damping by elastic fixation is a principal engineering strategy to increase the durability of load-bearing structures exposed to prolonged dynamic loading. This biomechanical study evaluated axial impact damping provided by a novel dynamic locking plate. In this design, locking screw holes are elastically suspended within a silicone envelope inside the locking plate. Axial impact damping was assessed for 3 distinct fixation constructs applied to bridge a 10-mm fracture gap of a femoral diaphysis surrogate: a standard locking plate, a dynamic locking plate, and an Ilizarov ring fixator. First, the 3 fixation constructs were characterized by determining their axial stiffness. Subsequently, constructs were subjected to a range of axial impact loads to quantify damping of force transmission. Compared with standard locked plating constructs, dynamic plating constructs were 58% less stiff (P < .01) and Ilizarov constructs were 88% less stiff (P < .01). Impact damping correlated inversely with construct stiffness. Compared with standard plating, dynamic plating constructs and Ilizarov constructs dampened the transmission of impact loads by up to 48% (P < .01) and 74% (P < .01), respectively. In conclusion, lower construct stiffness correlated with superior damping of axial impact loads. Dynamic locking plates provide significantly greater impact damping compared with standard locking plates.


Fundación Dialnet

Dialnet Plus

  • Más información sobre Dialnet Plus

Opciones de compartir

Opciones de entorno