Resumen de Characterization of long-term elution of platinum from carboplatin-impregnated calcium sulfate hemihydrate beads in vitro by two distinct sample collection methods

Rachel J. Tulipan, Heidi Phillips, Laura D. Garrett, Levent Dirikolu, Mark A. Mitchell

  • OBJECTIVE To characterize long-term elution of platinum from carboplatin-impregnated calcium sulfate hemihydrate (CI-CSH) beads in vitro by comparing 2 distinct sample collection methods designed to mimic 2 in vivo environments.

    SAMPLES 162 CI-CSH beads containing 4.6 mg of carboplatin (2.4 mg of platinum/bead).

    PROCEDURES For method 1, which mimicked an in vivo environment with rapid and complete fluid exchange, each of 3 plastic 10-mL conical tubes contained 3 CI-CSH beads and 5 mL of PBS solution. Eluent samples were obtained by evacuation of all fluid at 1, 2, 3, 6, 9, and 12 hours and 1, 2, 3, 6, 9, 12, 15, 18, 22, 26, and 30 days. Five milliliters of fresh PBS solution was then added to each tube. For method 2, which mimicked an in vivo environment with no fluid exchange, each of 51 tubes (ie, 3 tubes/17 sample collection times) contained 3 CI-CSH beads and 5 mL of PBS solution. Eluent samples were obtained from the assigned tubes for each time point. All samples were analyzed for platinum content by inductively coupled plasma–mass spectrometry.

    RESULTS Platinum was released from CI-CSH beads for 22 to 30 days. Significant differences were found in platinum concentration and percentage of platinum eluted from CI-CSH beads over time for each method. Platinum concentrations and elution percentages in method 2 samples were significantly higher than those of method 1 samples, except for the first hour measurements.

    CONCLUSIONS AND CLINICAL RELEVANCE Sample collection methods 1 and 2 may provide estimates of the minimum and maximum platinum release, respectively, from CI-CSH beads in vivo.