Effect of abutment cooling on reverse torque values of abutment screws: An in vitro study

  • Autores: Nirosa Srimurugan-Thayanithi, Samir Abou Ayash, Burak Yilmaz, Martin Schimmel, Urs Bragger
  • Localización: The International Journal of Oral & Maxillofacial Implants, ISSN-e 0882-2786, Vol. 38, Nº. 1, 2023, págs. 94-100
  • Idioma: inglés
  • Enlaces
  • Resumen
    • Purpose: To evaluate the effect of cooling on the reverse torque values of different abutments in bone-level and tissue-level implants. The null hypothesis was that there would be no difference in reverse torque values of abutment screws when cooled and uncooled implant abutments were compared.

      Materials and Methods: Bone-level and tissue-level implants (Straumann, each n = 36) were placed in synthetic bone blocks and subdivided into three groups (each n = 12) based on the abutment type (titanium base, cementable abutment, abutment for screw-retained restorations). All abutment screws were tightened to 35 Ncm torque. In half of the implants, a dry ice rod was applied on the abutments close to the implant-abutment connection for 60 seconds before untightening the abutment screw. The remaining implant-abutment pairs were not cooled. The maximum reverse torque values were recorded using a digital torque meter. The tightening and untightening procedure was repeated three times for each implant including cooling for the test groups, resulting in 18 reverse torque values per group. Two-way analysis of variance (ANOVA) was used to analyze the effect of cooling and abutment type on the measurements. Post hoc t tests were used to make group comparisons (α = .05). The P values of post hoc tests were corrected for multiple testing using the Bonferroni-Holm method.

      Results: The null hypothesis was rejected. Cooling and abutment type significantly affected the reverse torque values in bone-level implants (P = .004) but not in tissue-level implants (P = .051). The reverse torque values of bone-level implants significantly decreased after cooling (20.31 ± 2.55 Ncm vs 17.61 ± 2.49 Ncm). Overall mean reverse torque values were significantly higher in bonelevel implants compared to tissue-level implants (18.96 ± 2.84 Ncm vs 16.13 ± 3.17 Ncm; P < .001).

      Conclusion: Cooling of the implant abutment led to a significant decrease in reverse torque values in bone-level implants and may therefore be recommended as a pretreatment before the application of procedures to remove a stuck implant part.