Santiago, Chile
We report a 26year old patient who had a single chamber pacemaker implantation one year before. During a routine pre-operative evaluation, pacemaker dysfunction was demonstrated due to sensing and pacing failure, associated to left pectoral muscle rhythmic contraction. Chest X-ray confirmed Twiddler syndrome, in which twisting or rotation of the device inside the pocket results in lead dislodgement and device malfunction.
© 2001-2024 Fundación Dialnet · Todos los derechos reservados